banner
SYD-261D Fully Automatic Closed-Cup Flash Point Tester

SYD-261D Fully Automatic Closed-Cup Flash Point Tester

The Full automatic closed-cup flash point automatic tester is suitable for determining the flash point of petroleum products using the closed-cup method....

  • Fast Delievery
  • Quality Assurance
  • 24/7 Customer Service

SYD-261D Fully Automatic Closed-Cup Flash Point Tester


The Full automatic closed-cup flash point automatic tester is suitable for determining the flash point of petroleum products using the closed-cup method.


It is suitable for use in the petroleum, chemical, coating, railway, and aviation industries.


Meet standard

GB/T261、ASTM D93、ISO 2719


Features 蓝.jpg

1. 8-inch LCD touchscreen for real-time recording of experimental data and status parameters, with curve display; selectable A, B, and C testing methods.


2. Utilizes a 32-bit ARM processor, high-precision AD chip, high-precision temperature sensor, high-temperature igniter, imported motor, printer, and other core hardware components to ensure data accuracy and reliability.


3. Equipped with Bluetooth interface for unlimited cloud storage, allowing for data retrieval anytime, anywhere.


4. Remotely and automatically upgrades to the latest version, provides remote prediction and anomaly warnings, and comprehensively evaluates and maintains the instrument's operating status.


5. Automatic atmospheric pressure measurement and automatic correction of flash point values.


6. Automatic completion of heating, lid opening, ignition, testing, and data printing; automatic raising and lowering of the test arm; one-button completion of the entire flash point test process for a truly fully automated experience.


7. Electronic ignition, gas flame, automatic air cooling after the test, effectively improving experimental efficiency.


8. Available in both electronic and gas ignition modes to meet customer laboratory safety requirements.


Specifications.jpg

SYD-261D Full automatic Closed-cup Flash Point tester

Power supply

AC220±5%V 50Hz

Total power consumption

≤600W

Temperature range

RT230℃Resolution0.1℃

Sample temperature rising speed

Method A(56)℃/min
Method B(11.5)℃/min
Method C(3.0±0.5)℃/min

Stirring speed

⑴ (90120)r/min,suitable for Method A/B/C
⑵ (250±10)r/min,suitable for method B

Fire device

⑴ Piped gas or LPG
⑵ Electric fire(optional)

Communication

Bluetooth

Work environment

Temperature(1535)℃Humidity≤85%

Dimensions

510mm×345mm×330mmL*W*H

Net weight

20kg

261D.2.jpg