SYD-261-1 Closed-Cup Flash Point Tester

SYD-261-1 Closed-Cup Flash Point Tester

The Closed-Cup Flash Point Tester is a specialized instrument used to determine the flash point of petroleum products using the closed-cup method. ...

  • Fast Delievery
  • Quality Assurance
  • 24/7 Customer Service

SYD-261-1 Closed-Cup Flash Point Tester

The closed-cup flash point automatic tester is a specialized instrument used to determine the flash point of petroleum products using the closed-cup method. This involves heating the sample under specified conditions to the lowest temperature at which a mixture of its vapor and air ignites upon contact with a flame.


Applicable Standards: 

GB/T261, ASTM D93, ISO 2719


Features 蓝.jpg

1. Employs single-chip microcomputer control technology and an LCD screen display; features an English human-machine interface for easy operation.


2. The LCD screen displays experimental parameters and real-time sample temperature; when the flash point is detected, a simple press of the record button displays and saves the flash point value.


3. Utilizes a high-precision sensor to capture flash point values, and the system automatically records them; eliminating data deviations caused by manual observation, resulting in more accurate and reliable data.


4. Novel design and compact structure; users only need to connect it to coal gas or other domestic combustible gas for testing, making it convenient to use and providing accurate test results.


5. Equipped with a rapid cooling function to improve testing efficiency.


6. The instrument control program includes selectable methods A, B, and C, and the system automatically controls the heating rate and stirring speed, fully complying with GB/T 261 standard.


Specifications.jpg


SYD-261-1 Closed-Cup Flash Point Tester

Operating power supply

AC(220±10%)V 50Hz

Total power consumption

≤650W

Heating device

(1) The furnace body is made of silicon carbide material, and the heating wire has a power of 600W;

(2) The heating power is adjustable from (0600)W.

Sample heating rate

Step A(56)℃/min
Step B(11.5)℃/min
Step C(2.53.5) ℃/min

Electric stirring device

Stirring motor; Transmission method: flexible shaft connection; Stirring blade specifications: 8mm × 40mm

Stirring speed

1(90120)r/minsuitable for Step A
2(250±10)r/minsuitable for Step B
3(90120) r/minsuitable for Step C

Standard oil cup

1Inner diameter50.7mm50.8mm
2Depth55.7mm56.0mm
3Oil test capacity mark depth33.9mm34.3mm
4Oil test capacity:约70ml

Ignition gas source

Piped coal gas or liquefied petroleum gas.

Applicable environment

Temperature: (1535)℃, relative humidity: ≤85%.

Dimensions

340mm×330mm×380mmL*W*H

Net weight of the instrument

9kg